A clearer picture is not always more information.
Endoscopic image enhancement is now routine in ear, nose and throat practice, and the pictures on the screen look better than ever. Vessels that used to blend into pink mucosa stand out, and the deep corners of the sinuses and middle ear are evenly lit. Algorithms can sort a shaky throat video into frames worth reading and frames worth discarding.
The single word “enhancement” covers three kinds of technology, though. The first changes the light that reaches the tissue, so the camera records signals white light never collects. The second changes the pixels after the camera has recorded them, rearranging brightness and colour already in the image. The newest hands the video to an algorithm that discards poor frames or paints in what glare has hidden.
The three layers make different promises, which is why the distinction matters. A brighter, more colourful screen is easy to admire and hard to interrogate. This piece looks at how each layer works, what it makes easier to see, and how far it has moved from the laboratory into the clinic. It does not address whether any of these tools changes a diagnosis.
One thread runs through all three: the further down the list you go, the more of the screen is redrawn rather than newly captured.
Why the throat is a hard place to photograph
The larynx and pharynx are close to a worst case for a camera. The lining is wet and smooth, so it throws the endoscope’s own light straight back as white glare. The space is narrow and curved, the patient swallows and breathes, and during speech the vocal folds vibrate faster than the eye can follow. Light falls off quickly in the deeper airway, so some areas come out underexposed while others wash out.
Engineers who build quality-control software for laryngoscopy sort the resulting frames into four categories: informative, blurred, obscured by saliva or specular reflection, and underexposed [Moccia, Informative Laryngoscopic Frames, 2018]. Their first benchmark used 720 frames from 18 videos, balanced across those categories. The categories double as a map of the problem, since each technology below targets one or more of them.
How common are the bad frames? In one test of four recorded examinations, an AI quality checker judged no more than 60% of frames clear enough to read, which leaves at least four in ten unreadable [Baldini, Real-Time Informative Frames, 2024]. The sample is small and the verdict is a machine’s, but it matches the familiar clutter of a raw recording.

Anatomy sets the ceiling. A camera upgrade won’t remove a swallow, a cough or a film of mucus, and the later layers exist for that reason.
Layer one: changing the light (NBI and BLI)
The oldest and best-understood layer changes what lights the tissue. Olympus’s narrow band imaging (NBI) filters the endoscope’s light down to two narrow bands of blue and green, at 415 nm and 540 nm, that blood absorbs strongly [Masaki, NBI in Laryngeal SCC, 2009]. Because haemoglobin soaks up these wavelengths, blood vessels look dark against the surrounding mucosa. The shorter blue band is absorbed near the surface and reveals the finest capillaries; the green band reaches a little deeper and outlines larger vessels beneath them.
Researchers studied the optics systematically from the start. Early optical studies showed that the contrast of surface capillaries depends on both the centre wavelength and the width of the band, and that narrowing blue light around 415 nm sharpened the capillary pattern markedly [Gono, Narrow-Band Endoscopic Imaging, 2004]. The larynx and hypopharynx suit the method well: their thin, non-keratinised lining lets the surface vessel pattern show through [Piazza, NBI of the Larynx, 2012]. By 2006 the FDA had cleared an Olympus system whose light-source filter narrows white light into these bands (FDA, K061313, 2006).
This layer gathers new signal, because tissue lit differently reflects differently. Compare photographing a scene through a coloured filter with colour-correcting the photograph afterwards.

Small image studies measure the effect and agree on its direction. In 120 vocal folds with nodules, scars or sulci, NBI produced larger brightness changes across the lesion than ordinary light did [Pu, White Light vs NBI for Nodules, 2024]. In a single-centre study of 59 lesions in 30 patients from the pharynx and oesophagus, the colour difference between lesion and normal tissue was 11.6 under white light and 17.2 under NBI, about one and a half times the white-light gap [Dobashi, TXI Pharynx and Esophagus, 2021].
Even in the nose, where vessels are not the target, NBI separated cartilage from its covering membrane during septal surgery, with a brightness difference of 80.8 vs 35.6 (more than twice the contrast) across 26 sides in 15 patients [Yamamoto, NBI in Septoplasty, 2022].
Fujifilm took a different road to a similar place. Its blue light imaging (BLI) began on laser-based light sources and now runs on four independently controlled LEDs, first shown in Europe in 2016 (Fujifilm, ELUXEO Launch, 2016). BLI combines short-wavelength light with image processing to emphasise surface vessels and structure [Fujifilm, ELUXEO 8000 Release, 2024].
The limits of a better light
Narrow-band light is dimmer than white light, which the vocal-fold images above confirm directly [Pu, White Light vs NBI for Nodules, 2024]. The picture also fades with distance. In a comparison that included pharyngeal lesions, only a purpose-built “bright” BLI mode held its brightness and contrast at a working distance of about four centimetres (40 mm); standard narrow-band modes faded [Kaneko, Bright BLI, 2014]. That test used an older laser-based system, but the trade-off is general: narrower light means less light.
The mechanism sets a second limit. NBI reads surface vessels, so it says little about growths beneath an intact lining, and inflammation or radiation change can produce vessel patterns that resemble more worrying ones [Valls-Mateus, NBI Lessons from Mistakes, 2018]. The method shows the vessels it can reach and nothing beneath them.
Layer two: changing the pixels (SPIES, TXI and LCI)
The second layer leaves the light alone and works on the image the camera has already recorded. KARL STORZ’s Professional Image Enhancement System (SPIES) is the pixel-layer system that appears most often in the ENT studies reviewed here. Its CLARA mode brightens dark areas to even out illumination, CHROMA strengthens contrast and colour, and the SPECTRA modes shift the colour spectrum to emphasise selected parts of it [Zhou, IMAGE 1 SPIES, 2016].
Storz describes these modes as real-time processing of the camera image (KARL STORZ, IMAGE1 S, 2026), and researchers classify Spectra A and B as “virtual chromoendoscopy,” meaning colour separation done in software rather than with a different light [Huang, SPIES Image Biomarkers, 2026]. Spectra A draws mainly on the blue and green signals; Spectra B shifts colour tone to suppress the dominant red reflection.
Olympus offers the same idea in its newer platform. Texture and colour enhancement imaging (TXI), introduced with the EVIS X1 system in 2020, splits the picture into a brightness layer and a detail layer, lifts dark areas, sharpens texture and, in one mode, boosts colour [Olympus, EVIS X1 Release, 2020]. The same platform includes BAI-MAC, which brightens dark regions while preserving texture, much like CLARA (Olympus, EVIS X1, 2026). Fujifilm’s linked colour imaging (LCI) also belongs largely here: it emphasises slight colour differences in the red range of the image [Fujifilm, ELUXEO 8000 Release, 2024].

Surgeons like these tools. In one sinus-surgery series of 27 patients and 10 healthy volunteers, surgeons using SPIES reported clearer views of mucosa and vessels, no added time and no distracting distortion [Englhard, SPIES in Sinus Disease, 2021]. A panel of 51 ear surgeons rated CLARA, the brightening mode, as the most useful for cholesteatoma surgery, ahead of the two SPECTRA modes [Ragonesi, Digital Enhancement in Ear Surgery, 2023].
For TXI, the closest evidence is one small image study of the pharynx and oesophagus. In its colour mode, the lesion-to-mucosa colour difference was 18.6, roughly 60% larger than under white light, and most reviewers judged 37 of 59 lesions easier to see. About six in ten lesions looked clearer, and none looked worse [Dobashi, TXI Pharynx and Esophagus, 2021]. These are impressions and image measurements, but they point the same way.
| Technology | What it changes | What becomes easier to see | Main home | Introduced |
|---|---|---|---|---|
| NBI (Olympus) | The light: two narrow blue-green bands | Surface capillaries and vessel patterns | Larynx, pharynx; also GI | 2006 (US clearance) |
| BLI (Fujifilm) | Light plus processing: short-wavelength emphasis | Surface vessels and fine structure | Mainly GI; throat seen during GI endoscopy | 2016 (LED platform, Europe) |
| SPIES CLARA / CHROMA (Storz) | Pixels: brightness and contrast | Dark corners of sinus and middle ear; mucosal contrast | ENT surgery | 2014 (US announcement) |
| SPIES SPECTRA A / B (Storz) | Pixels: colour separation and shift | Vessel-rich mucosa, red-dominated scenes | ENT surgery | 2014 (US announcement) |
| TXI and BAI-MAC (Olympus) | Pixels: brightness, texture and colour layers | Subtle colour and surface changes; dark areas | Mainly GI and airway | 2020 |
| LCI (Fujifilm) | Mostly pixels: red-range colour difference | Inflammation and subtle redness | Mainly GI | 2016 (LED platform, Europe) |
Where processing stops helping
More processing doesn’t automatically help. In a urology study, the colour-shift SPECTRA modes were rated below plain white light, while CLARA was rated above it [Emiliani, SPIES Image Quality, 2017]. In the same pharynx-and-oesophagus study, the texture-only mode gave a smaller boost: a colour gap of 14.3, about a quarter larger than under white light, with less gain in visibility than the colour mode [Dobashi, TXI Pharynx and Esophagus, 2021].
The deeper limit is conceptual. Pixel-level tools cannot add a signal the camera never recorded. They rearrange contrast already in the image, which can make a real difference easier to notice and can also make an unremarkable one look emphatic. They improve the lighting of one photograph; they can’t take another.
Layer three: letting algorithms clean the video
The newest layer treats the endoscopy video as data. Its most mature task is triage: deciding, frame by frame, which images deserve a clinician’s attention. After the first benchmark of 720 frames from 18 videos [Moccia, Informative Laryngoscopic Frames, 2018], later work with pretrained networks improved recognition of good frames [Patrini, Transfer Learning Frame Selection, 2020].
In a study using 5,147 internal and 646 external images, a deep-learning model sorted white-light and NBI frames in real time with a precision of 95% (internal) and 97% (external), so roughly 19 of every 20 frames it called usable were usable [Baldini, Real-Time Informative Frames, 2024]. Its recall of 97% (internal) and 89% (external) tells the other half: on another hospital’s images it missed about one usable frame in nine (11%).
These results are early but consistent. They come from research prototypes tested on recorded videos, and the authors present the system as a step toward clinical development, not a finished product. Regulators keep a running list of authorised AI devices (FDA, AI-Enabled Medical Devices, 2026); the throat-video triage tools described here are research systems.
Removing glare is a more ambitious task. Several groups detect the bright reflections and fill them with plausible tissue, using large pretrained inpainting models [Li, EndoSRR, 2024] or networks that borrow information from neighbouring video frames where the glare has moved [Daher, Inpainting Endoscopic Specularities, 2023]. In surgical and stomach video these tools make images look cleaner, and surgeons have scored the results favourably [Zhang, Specular Highlight Removal, 2023]. The literature reviewed for this piece contained no test on throat endoscopy, and what the tools paint in is the algorithm’s best guess.
High-speed video: the light budget
High-speed videoendoscopy, used to study vocal-fold vibration, shows why this layer exists. A camera shooting thousands of frames a second gives each frame a sliver of light, so camera sensitivity is a central constraint. A widely used rate of 2,000 frames per second may be too slow for some vibration features, while the best cameras of the time reached 8,000 frames per second in colour and 20,000 in monochrome. The most sensitive sensors offered 7,000 ISO in monochrome against 2,100 ISO in colour, so colour cameras had about a third of the light sensitivity [Deliyski, Laryngeal Imaging, 2010]. The figures are more than a decade old; the trade-off between speed and light still holds.
So far, AI mostly helps trace the vocal-fold edges in these dim, fast frames rather than clean them up. A shared benchmark of 59,250 frames from 640 recordings at seven institutions let researchers train and compare such models [Gómez, BAGLS, 2020], and one network reached an intersection over union of 0.81 (its traced outline overlapped the expert’s by about four-fifths) even in poor-quality frames [Yousef, Vocal Fold Dynamics HSV, 2022]. Real-time noise removal for laryngeal high-speed video would be a natural next step, but this review found no validated example.
When endoscopic image enhancement can mislead
The first two layers adjust a picture of something the camera recorded. Generative repair works differently. When an algorithm fills in what the camera could not record, it draws on patterns learned from other images, and it can draw things that are not there. Imaging scientists call these hallucinations: an inaccurate learned prior can place false structures into a reconstructed image [Bhadra, Hallucinations in Reconstruction, 2021].
This has already happened. In a recent CT study, an AI super-resolution tool generated 16 of 80 nodules (20%, one in five) that the authors classed as hallucinated, mostly benign or artifactual structures [Kang, CT Super-Resolution Nodules, 2026]. CT is a different kind of scan, but the risk is closely related. Glare-removal networks that infer hidden anatomy from neighbouring frames are, by design, reporting a guess about tissue the camera never saw [Daher, Inpainting Endoscopic Specularities, 2023].
The earlier layers have quieter versions of the same problem. Narrow-band images darken with distance, and inflammation can imitate surface-vessel patterns [Valls-Mateus, NBI Lessons from Mistakes, 2018]. Colour-shift modes can look more dramatic without showing more [Emiliani, SPIES Image Quality, 2017]. The tools are still useful, but the viewer needs to know which layer produced what is on the screen.

Adoption follows the same gradient. In the ENT literature, NBI and, more recently, SPIES dominate. Fujifilm’s and Olympus’s newer modes were built for gastroenterology, so they usually meet the throat on the way down to the stomach [Iwamuro, Pharyngolaryngeal Papillomas, 2025]. AI frame triage and glare repair remain largely in research.
Every enhancement shows something. The question is whether it was there to be seen.
Key Takeaways
- Endoscopic image enhancement in ENT works on three distinct layers: the light, the pixel and the algorithm.
- Narrow band imaging gathers new signal by lighting tissue with wavelengths that blood absorbs, so surface vessels appear dark and sharp.
- Digital modes such as SPIES and TXI rearrange brightness and colour already in the image. Surgeons report them most useful for evening out light in deep spaces like the sinuses and middle ear.
- AI frame triage can sort usable from unusable laryngoscopy frames in real time, but published tests so far have used recorded videos, not routine clinics.
- Generative glare removal fills gaps with the algorithm’s best guess, and similar tools in other imaging fields have produced structures that do not exist.
- Trust an enhanced image in proportion to how much of it was captured rather than redrawn.
FAQ
Is narrow band imaging just a colour filter applied to the video?
No. NBI changes the light that reaches the tissue, narrowing it to bands at about 415 nm and 540 nm that blood absorbs, so the camera records a vessel pattern white light does not produce [Masaki, NBI in Laryngeal SCC, 2009]. Digital modes that colour-correct an existing image work differently.
Does SPIES use a different light source from ordinary endoscopy?
No. Storz describes its CLARA, CHROMA and SPECTRA modes as real-time processing of the camera image, and researchers classify the SPECTRA modes as virtual chromoendoscopy done in software [Huang, SPIES Image Biomarkers, 2026]. The light is ordinary white light; the processing changes how the image is displayed.
Can AI remove glare from throat endoscopy videos?
Not in routine practice yet. Glare-removal networks have been tested on surgical and stomach video, where they fill reflections with inferred tissue [Daher, Inpainting Endoscopic Specularities, 2023]. This review found no published test in throat endoscopy, and the filled areas are estimates, not recordings.
Why are so many laryngoscopy frames unusable?
Motion, glare and darkness are constant problems in the throat. In one test of four recorded examinations, an AI checker judged at least four in ten frames not clear enough to read [Baldini, Real-Time Informative Frames, 2024]. Swallowing, saliva and reflections from wet mucosa all contribute.
References
- Moccia S, Vanone GO, De Momi E, et al. Learning-based classification of informative laryngoscopic frames. Comput Methods Programs Biomed. 2018;158:21-30.
- Baldini C, Azam MA, Sampieri C, et al. An automated approach for real-time informative frames classification in laryngeal endoscopy using deep learning. Eur Arch Otorhinolaryngol. 2024;281(8):4255-4264.
- Patrini I, Ruperti M, Moccia S, Mattos LS, Frontoni E, De Momi E. Transfer learning for informative-frame selection in laryngoscopic videos through learned features. Med Biol Eng Comput. 2020;58(6):1225-1238.
- Gono K, Obi T, Yamaguchi M, et al. Appearance of enhanced tissue features in narrow-band endoscopic imaging. J Biomed Opt. 2004;9(3):568-577.
- Masaki T, Katada C, Nakayama M, et al. Narrow band imaging in the diagnosis of intra-epithelial and invasive laryngeal squamous cell carcinoma: a preliminary report of two cases. Auris Nasus Larynx. 2009;36(6):712-716.
- Piazza C, Del Bon F, Peretti G, Nicolai P. Narrow band imaging in endoscopic evaluation of the larynx. Curr Opin Otolaryngol Head Neck Surg. 2012;20(6):472-476.
- Pu S, Laitman B, Woo P. Objective comparison of white light and narrow-band imaging for detecting scars, sulci and nodules. Laryngoscope. 2024;134(9):4066-4070.
- Dobashi A, Ono S, Furuhashi H, et al. Texture and color enhancement imaging increases color changes and improves visibility for squamous cell carcinoma suspicious lesions in the pharynx and esophagus. Diagnostics (Basel). 2021;11(11):1971.
- Yamamoto K, Okuni T, Kurose M, et al. Narrow band imaging accentuates differences in contrast between cartilage and perichondrium in the elevation of the muco-perichondrium flap during septoplasty and open septorhinoplasty. Auris Nasus Larynx. 2022;49(6):973-979.
- Kaneko K, Oono Y, Yano T, et al. Effect of novel bright image enhanced endoscopy using blue laser imaging (BLI). Endosc Int Open. 2014;2(4):E212-E219.
- Valls-Mateus M, Nogués-Sabaté A, Blanch JL, Bernal-Sprekelsen M, Avilés-Jurado FX, Vilaseca I. Narrow band imaging for head and neck malignancies: lessons learned from mistakes. Head Neck. 2018;40(6):1164-1173.
- Zhou L. IMAGE 1 SPIES in urology. Transl Androl Urol. 2016;5(Suppl 1):AB019.
- Huang Z, et al. Artificial intelligence-optimized image biomarker identification for sinonasal inverted papilloma diagnosis using image-enhanced versus white-light endoscopy. World J Otorhinolaryngol Head Neck Surg. 2026.
- Englhard AS, Ledderose C, Volgger V, Ledderose GJ. Evaluation of an image enhancement system for the assessment of nasal and paranasal sinus diseases. Am J Otolaryngol. 2022;43(2):103323. Epub 2021 Dec 14.
- Ragonesi T, Niederhauser L, Fernandez IJ, et al. Digital image enhancement may improve sensitivity of cholesteatoma detection during endoscopic ear surgery. Clin Otolaryngol. 2023;48(4):595-603.
- Emiliani E, Talso M, Baghdadi M, et al. Evaluation of the Spies modalities image quality. Int Braz J Urol. 2017;43(3):476-480.
- Li W, Jia F, Liu W. EndoSRR: a comprehensive multi-stage approach for endoscopic specular reflection removal. Int J Comput Assist Radiol Surg. 2024;19(6):1203-1211.
- Daher R, Vasconcelos F, Stoyanov D. A temporal learning approach to inpainting endoscopic specularities and its effect on image correspondence. Med Image Anal. 2023;90:102994.
- Zhang C, Liu Y, Wang K, Tian J. Specular highlight removal for endoscopic images using partial attention network. Phys Med Biol. 2023;68(22).
- Deliyski DD, Hillman RE. State of the art laryngeal imaging: research and clinical implications. Curr Opin Otolaryngol Head Neck Surg. 2010;18(3):147-152.
- Gómez P, Kist AM, Schlegel P, et al. BAGLS, a multihospital Benchmark for Automatic Glottis Segmentation. Sci Data. 2020;7(1):186.
- Yousef AM, Deliyski DD, Zacharias SRC, Naghibolhosseini M. Deep-learning-based representation of vocal fold dynamics in adductor spasmodic dysphonia during connected speech in high-speed videoendoscopy. J Voice. 2025;39(2):570.e1-570.e15. Epub 2022 Sep 23.
- Bhadra S, Kelkar VA, Brooks FJ, Anastasio MA. On hallucinations in tomographic image reconstruction. IEEE Trans Med Imaging. 2021;40(11):3249-3260.
- Kang SR, Jeong WG, Rho JY, et al. Impact of CT slice thickness reduction algorithm on AI-based lung nodule detection in chest CTs of colorectal cancer patients. Quant Imaging Med Surg. 2026;16(9):740.
- Iwamuro M, et al. Endoscopic features of oral and pharyngolaryngeal papillomas and their role in distinguishing squamous cell carcinoma. World J Gastrointest Endosc. 2025;17(12):110594.
- FDA (US Food and Drug Administration). 510(k) premarket notification K061313: EVIS EXERA II 180 system. Olympus Medical Systems Corp.; 2006 Aug 30. https://www.accessdata.fda.gov/cdrh_docs/pdf6/K061313.pdf
- Fujifilm Europe. Fujifilm launches ELUXEO, the next generation of endoscopy systems featuring a unique 4-LED Multi Light technology [press release]. PR Newswire; 2016 Oct 19.
- Fujifilm Corporation. ELUXEO 8000 system launch [press release, Japanese]. 2024 May 28. https://www.fujifilm.com/jp/ja/news/list/11398
- Olympus Corporation. EVIS X1 launch in Japan [press release, Japanese]. 2020 Jul 1. https://www.olympus.co.jp/news/2020/contents/nr01704/nr01704_00002.pdf
- Olympus. EVIS X1 product information. https://www.olympus.com.au/medical/en/Products-and-Solutions/Products/Product/EVIS-X1-Bronchoscopy.html
- KARL STORZ. IMAGE1 S product insights. https://insights.karlstorz.com/KSUS-IMAGE1S
- FDA (US Food and Drug Administration). Artificial intelligence-enabled medical devices. Updated 2026 Sep 4. https://www.fda.gov/medical-devices/software-medical-device-samd/artificial-intelligence-enabled-medical-devices
Joonpyo Hong, MD is a board-certified otolaryngologist practicing in Korea. This article reflects his clinical interpretation of published research and does not constitute individual medical advice. This article is not intended to advertise or promote any specific company or product.
For more articles:
https://curiousmd.com/cnn-laryngeal-cancer-diagnosis/
https://curiousmd.com/convolutional-neural-networks-in-medicine/
Link out to:
FDA 510(k) clearance letter for an Olympus endoscopy system with narrow band imaging (2006) — US Food and Drug Administration
Artificial Intelligence-Enabled Medical Devices list — US Food and Drug Administration
EVIS X1: TXI, RDI, NBI and BAI-MAC imaging modes — Olympus
ELUXEO 8000: LED multi-light, BLI, LCI and ACI — FUJIFILM Healthcare Americas
IMAGE1 S camera platform and Clara + Chroma processing — KARL STORZ
2014 announcement of the IMAGE1 SPIES visualization tools — KARL STORZ Endoscopy-America via BioSpace